Vous êtes :
The INFINEED lt;Gravity lt; 3-ways infusion set without DEHP, BPA-free was designed to minimize the environmental impact of the device. Two methods have been used: first, by reducing the plastic mass used by around 10%, without making any changes to functionality; second, by exclusively using raw materials that are free from endocrine disruptors (elimination of components comprising DEHP or BPA.)
The polycarbonate-free 3-way stopcock has been tested at pressures well above those recommended for gravity infusion; laboratory tests have shown a pressure resistance of approximately 10 bar.
Infusion set with:
- Two-channel perforator with grip wings
- Air intake: with 0.2µm filter. Delivered closed for immediate use on infusion bags.
- Chamber: Soft and translucent, the chamber is calibrated at 20 drops = 1ml ± 0.1ml. 15µ particle filter.
- Flow regulator: ABS linear flow control for good roller clamp rigidity, precision and reliability.
- BPA-free 3-way stopcock: Lipid-resistant with large valve stopcock and arrows indicating the direction of flow, permanently attached to the extension tube. Only the main line can be disconnected from the stopcock.
- Tubing: PVC (DINCH) designed for comfort, optimal flow adjustment consistency and to prevent kinks. Diameter: 3 x 4.1mm
- End connector: Mobile Luer Lock
- End stopper: 0.8 µm purge filter with hydrophobic membrane
Length: 150cm+60cm
Residual volume: 18.9 ml
Made in China.
Contact with 0.5% or 2% Chlorhexidine in 70° Isopropyl alcohol, without evaporation time, and while the luer lock is connected, therefore under physical stress, presents a significant risk of cracking of the luer lock, and should be avoided as much as possible.
In accordance with the recommendations of the SF2H (2019) and Didactic's disinfection protocol, it is always preferable to disinfect the connectors (when necessary) with 70° ethyl alcohol followed by a mandatory drying time of 15 secondes at least before connecting. 70° ethyl alcohol is just as effective disinfecting inert materials and then reduces the risk of altering plastic materials.
Indications
This device enables the parenteral administration by gravity of injectable preparations: parenteral drugs, solutions and nutrition.
For peripheral lines, the infusion line can remain in place for a maximum of 96 hours (i.e. the maximum time the peripheral catheter can be in place).
For central lines, the infusion line must be changed at least every 72 hours (or according to the institution's validated protocol).
Dispose according to regulation in member state.
Matériaux utilisés
PVC DINCH (DEHP<0.1%(m/>)): Chamber, air intake, tubing - Polypropylene: Protective cap, stopcock stopper - Polyethylene: Stopcock valve, purge filter - Tritan® BPA-free: Stopcock body - Acrylonitrile butadiene styrene: Perforator, regulator, mobile Luer Lock - Nylon: Particle filter
Étiquetage
on each unit are written the reference, the lot/serial number, the sterilization method and the expiry date if applicable
dispositif médical
stérile
conformité
opérateur économique
Conditions de transport
Protect from light and humidity and store at a temperature between 5°C and 20°C.
| Conditionnement | Quantités de pièces | Dimensions en cm | Poids en kg | Ean 13 du conditionnement |
| UNIT | 1 | 10 x 25 x 1 | 0.031 | 3661809029394 |
| BAG | 25 | 48 x 29 x 5 | 0.840 | 3661809129391 |
| CASE | 250 | 58 x 40 x 25 | 8.000 | 3661809229398 |